Hoefer TE22 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Ekwipunek Hoefer TE22. Hoefer TE22 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Hoefer TE22

user manualum TE22-IM/Rev.G0/07-12Hoefer TE22Tank transfer unit

Strona 2 - Contents

• p2SpecificationsGel size up to four 9 × 10 cm gelsMax. wattage 50 WMax. voltage 100 VMax. amperage 500 mAMax. temperature 45 °CBuffer required

Strona 3 - Vigtig Information – Danish

• p3Included but not shown:• Gel cassettes (4)• Foam sponges, 6 mm thick (4)• Foam sponges, 3 mm thick (8)• Blotting paper, sheets (25)coverelectrode

Strona 4 - Tärkeää Tietoa – Finnish

• p4Operating instructionsPerform the transfer as soon as possible after electrophoresis to minimize band diffusion. Each step is described below. Pre

Strona 5

• p5Prepare two lengths of 9 mm (3/8") vinyl or silicone tubing. Slide hose clamps (4 total) onto each end of two lengths of tubing. Attach one e

Strona 6 - Portuguese

• p6sponge, and then place the membrane on the blot-ting paper. Place the gel—which contains a sample that has been electrophoretically separated and

Strona 7 - Viktig Information – Swedish

• p7Install the cassette(s)1If transferring only one or two gels, choose the cassette positions nearest the center. The cassettes must be oriented so

Strona 8 - Electronic Equipment (WEEE)

• p8Final assembly and transfer1Install the lidThe cassettes are color coded to match the leads in the lid. To transfer toward the anode, orient the l

Strona 9 - Unpacking

• p94Set the power supplyConstant current mode is recommended. If constant voltage mode is selected, carefully monitor the current (increased current

Strona 10 - Specifications

• p10Care and maintenanceCleaning• Do not autoclave or heat any part above 45 °C.• Do not expose to alcohols or organic solvents!*• Never use abrasiv

Strona 11

• p11Troubleshootingproblem solutionIncomplete transfer Blank areas on Remove all trapped air pockets in the transfer stack assembly: the membr

Strona 12 - Operating instructions

• piContents Important Information ... iiWaste Electrical and Electronic Equipment (WEEE) ...

Strona 13

• p12problem solutionDiffuse band patterns Transfer immediately after electrophoretic separation. If equilibrat-ing before the transfer, shorten or

Strona 14

• p13Electrotransfer notesElectrophoretic transfer advantagesElectrophoretic transfer of proteins and nucleic acids is much faster than the blotting m

Strona 15 - Install the cassette(s)

• p14Factors affecting the transferParameters such as sample characteristics, membrane type, gel pore size, and the transfer buffer used all contribut

Strona 16 - Final assembly and transfer

• p15Instrument guidelinesCoolingConsiderable Joule heat is generated during any transfer because of the high current employed, so active cooling is r

Strona 17

• p16Protein transfersStudy summariesGershoni and Palade (1982) investigated factors affecting protein recovery from SDS gels to nitrocellulose or DBM

Strona 18 - Care and maintenance

• p17Protein transfer buffersUse a buffer with low ionic strength, such as the two listed below, to prevent overheating. Use the alternate CAPS buffer

Strona 19 - Troubleshooting

• p18Nucleic acid transfersNucleic acids must normally be transferred in denatured form for most efficient binding. RNA is normally denatured with gly

Strona 20

• p19EDTA solutiona(0.5 M EDTA, pH 8.0, 100 ml)Na2EDTA·2H2O (FW 372.2) 0.5 M 18.6 gDissolve in 70 ml distilled water. Adjust to pH 8.0 with 10 M NaO

Strona 21 - Electrotransfer notes

• p20BibliographyAlwine, J.C., Kemp, D.J., and Stark G.R., Method for detection of specific RNAs in agarose gels by trans-fer to DBM paper and hybridi

Strona 22

• p21Lin, W., and Kasamatsu, H., On the electrotrans-fer of polypeptides from gels to nitrocellulose membranes. Anal. Biochem. 128, 302–311 (1983).Mat

Strona 23 - Instrument guidelines

• piiImportant Information – English• If this equipment is used in a manner not speci-fied by Hoefer, Inc. the protection provided by the equipment m

Strona 24 - Protein transfers

• p22Ordering informationproduct quantity code no.Hoefer TE22 Tank Transfer Unit. 1 TE22 Includes: 4 gel cassettes, 8 foam sponges, 3-mm thi

Strona 25 - Protein transfer buffers

• p23Hoefer, Inc. 84 October Hill Road Holliston, MA 01746Toll Free: 1-800-227-4750 Phone: 1-508-893-8999 Fax: 1-508-893-0176 E-mail: support@hoeferin

Strona 26 - Nucleic acid transfers

• piiiBelangrijke Informatie – Dutch• Indien deze uitrusting in een manier wordt gebruikt die niet door Hoefer, Inc. is gespecificeerd de bescherming

Strona 27 - RNA transfer buffer, 10X

• pivl’instrument. Les dissolvants organiques causeront des dommages irréparables à l’unité!• Ne pas fonctionner avec les températures de tampon au-d

Strona 28 - Bibliography

• pv• Sikkerheten lokket må være på plass før forbinding kraftforsyningene blyene til en kraftforsyning. • Vender all kraftforsyningsstyring av og f

Strona 29

• pvipor Hoefer, Inc. puede ser utilizado para operar, para mantener, y para atender a este producto.• Sólo utiliza una alimentación que es CE marcó

Strona 30 - Ordering information

• pviiEnglishFrenchGermanItalianSpanishSwedishThis symbol indicates that the waste of electrical and electronic equipment must not be disposed as unso

Strona 31

• p1Transfer Electrophoresis Unit function and descriptionThe Hoefer® TE22 Tank Transfer unit rapidly transfers proteins, DNA, or RNA from up to four

Komentarze do niniejszej Instrukcji